Micronics U1000MKII-FM | Ultrasonic flow meter

Contact Instrument Solutions

The U1000MKII-FM is an ultrasonic permanent/fixed clamp-on flow metering solution for measuring flow rate and total flow with a volume pulse output and optional Modbus or 4-20 mA flow proportional output, which can be used as a stand alone meter or as part of an integral management system.

The instrument is simple to install by connecting power and entering the pipe inside diameter, adjust the sensors and clamping-on to the pipe with no specialist skills or tools required. It is a cost effective alternative to traditional in-line meter installation and with dry servicing it provides minimum downtime and maximum availability.

Compact, rugged and reliable, the U1000MKII-FM has been designed to provide sustained performance in industrial environments.

PLEASE NOTE - The U1000MKII-FM is available in several options, please contact a member of the Instrument Solutions sales team to discuss your requirements.

Features

  • Larger pipe range with 2 options, fully adjustable for 22 mm – 115 mm and 125 mm – 180 mm OD
  • Optional Modbus RTU slave, RS485 serial
  • Cross correlation flow measurement system
  • 1 Mhz transducers to get through poor/old thick wall pipes and larger pipework
  • Water/Glycol option
  • Simple to install: connect power and enter the pipe inside diameter, adjust the sensors and clamp-on the pipe, no specialist skills or tools required
  • A clamp-on non-invasive alternative to traditional in-line meter installation, plus dry servicing, providing minimum downtime and maximum availability.
  • LCD with backlight – install & service information.
  • Integrated pulse and optional 4-20 mA output – Automatic Monitoring & Targeting System (aM&TS) or Building Energy Management system (BEMS) compatible with Modbus.

Technical

Accuracy      +/- 1 % – 3 % of flow reading for > 0.3 m/s

Flow velocity range      0.1 to 10 m/s, bidirectional

Pipe range       Available in 2 options, fully adjustable for 22 mm – 115 mm and 125 mm – 180 mm OD

Pipe material      Steel, stainless steel, plastic and copper

Water temperature range      0 °C – 85 °C

4 - 20 mA output      Optional 4 - 20 mA flow proportional output

Pulse output Pulse or frequency. Pulse for volume flow and alarms. Frequency for flow rate. The pulse output can be configured as loss of signal or low flow alarm

Modbus communication       Optional Modbus RTU slave, RS485 serial link hardware layer. Connection cable 1 M

External power supply      12 V - 24 V +/- 10 % AC/DC at 7 watts. Optional plug in 12 V power supply

Electronics enclosure      IP54

Input/output cable       5 m x 6 core for power in, 4 - 20 mA and pulse output

Micronics U1000MKII-HM | Ultrasonic heat meter
Micronics U1000MKII-HM | Ultrasonic heat meter
Available to purchase for pipes up to 180 mm diameter is the Micronics Ultraflo U1000MKII-HM, an ultrasonic clamp-on heat/energy meter that both displays and outputs the measured parameters

Support documents

  • Download
  • Category
  • File type
  • Download
Download
Micronics U1000MKII FM user manual.pdf Micronics U1000MKII FM user manual
Category
Technical document
File type
Adobe Acrobat
Download (1 MB)