The Testo 175T series of data loggers are small and portable which makes them the ideal device for monitoring temperatures in the modern workplace. Typical applications include the commissioning of buildings, the validation of temperature cold rooms and the monitoring of freezer temperatures. Logger size 82mm (H) x 52mm (W) x 30mm (D).
Features
General logger hire information
All hire data loggers are set as a default to record at 5-minute intervals with a user manual start on site by pressing and holding GO for 4-5 seconds. If an alternative logging interval or a timed start is required, then we must be notified of this prior to the despatch of the hire instruments.
For the hire of 1-9 units we do not charge for the time to download the instruments when they are returned to our Bracknell offices and will supply the data in Excel spread sheet format via email. For the hire of 10 or more units the costs for the data download is charged at an hourly rate. If downloaded data is required in a graphical format additional charges are also applied based on specific client requirements.
The 175T3 instruments have adjustable logging intervals via the software from 10 seconds to 24 hour increments, with a 16,000 sized readings memory. This for example equates to:-
16000 size memory (2 channels)
30 second logging interval = 2.8 days maximum capacity
1 minute logging interval = 5.6 days maximum capacity
5 minute logging interval = 27.8 days maximum capacity
10 minute logging interval = 55.6 days maximum capacity
NOTES: If x1 channel is switched off the above periods can be doubled.
All probes and extension leads for use with the 175T3 product are charged for at an additional cost.
Technical
175T3 specification (2 channel temperature logger with external connections for Type T or Type K sensors)
Storage temperature -40 to +85°C
Operating temperature 0 to +85°C
Memory 16 000 readings
Probe input Type T
Measuring range -50 to +400 °C
Accuracy ±0.5 °C (-50 to +70 °C) ±0.7 % of mv (+70.1 to +1000 °C)
Resolution 0.1 °C
Probe input Type K
Measuring range -50 to +1000 °C
Accuracy ±0.5 °C (-50 to +70 °C) ±0.7 % of mv (+70.1 to +1000 °C)
Resolution 0.1 °C