Chauvin Arnoux L452 | Voltage and current data logger

Contact Instrument Solutions

Ideal for industrial process applications such as the outputs from various types of environmental sensor such temperature, humidity and air velocity. The Chauvin Arnoux L452 is equipped with 2 input using 4 to 20 mA and 0 to 10 V modules. For each of the channels, a real-time measurement is displayed on the LCD screen. The L452 communicates via a USB input and the Bluetooth function for configuration and data transfer onto a PC. The unit's ergonomics make it simple to use including magnetized rear panel, Multifix system and desktop stand. It operates on rechargeable batteries or the mains supply.

Features

  • Oscilloscope, multimeter, analyser and recorder modes, with the recorded files directly viewable on the oscilloscope
  • Safety features include channels genuinely isolated from one another and from the earth, 6 V CAT III & Probix probes
  • Control panel for the data analysis software is a module that is fully integrated in the Dataview software
  • The instrument can also be configured using its front panel, with a
    very simple user interface
  • Multilingual menu options
  • Alarms can be set on the various measurements
  • Counter mode displays the measurement results in the unit selected
  • Programmable measurement campaign can be launched simply using the L452’s keypad
  • Users can lock the instruments keypad with the PC software

Technical

Battery life      200 ms acquisition, display on: 18 days, 200 ms acquisition, display off: 36 days and 1 min. acquisition, display off: 270 days

Power supply       110 to 240 V (50/60 Hz) – External: via USB connector – Internal: rechargeable NiMH batteries, 2.4 V (2 x 1.2 V)

Recording modes       Start/Stop (stop when memory full or when campaign end date is reached)

Control       Local mode (multi-directional keypad on front panel) – Remote mode (control via PC)

Recording duration      10 minutes to 1 year, modifiable e.g. 2 channels at @ 200 ms: 19 days and 2 channels @ 1 min: > 1 year (theoretically)

Acquisition interval       200 ms to 1 hour

Communication       Bluetooth 2.1, Class 1, USB 2.0

Dimensions       32.4 mm x 65.5 mm x 125 mm (137.5 mm with screw connector)

Weight       206 g

Display       LCD 128 x 64 pixels

Measurement terminal strip      6 screw terminals

Temperature       Operation: 0 to 50 °C

Protection       IP42 (IP20 terminal)

Electrical protection      IEC 61010-1 Ed. 3 and IEC 61010-2-030 Ed. 1

DC current:

Measurement range       4 to 20 mA

Resolution       0.01 mA

Accuracy (% of reading)        0.05 mA (0.25 %)

Sampling rate      5 S/s

DC voltage:

Measurement range       ± 100 mV, ± 1 V and ± 10 V

Resolution       0.1 mV, 1 mV and 10 mV

Accuracy (% of reading)        ± 0.1 mV (± 0.5 %), ± 1 mV (± 0.5 %) and ± 10 mV (± 0.5 %)

Sampling rate     5 S/s

Pulse:

Resolution      1 ms

Digital:

Resolution      1 ms

Accuracy (% of reading)     1 s (recording for 1 month max.)

Pulse voltage      3.3 V (with 1,000,000 Ω pull-up)

Support documents

  • Download
  • Category
  • File type
  • Download
Download
Chauvin Arnoux L452 data sheet.pdf Chauvin Arnoux L452 data sheet
Category
Technical document
File type
Adobe Acrobat
Download (2 MB)
Download
Chauvin Arnoux L452 quick start guide.pdf Chauvin Arnoux L452 quick start guide
Category
Technical document
File type
Adobe Acrobat
Download (6 MB)
Download
Chauvin Arnoux L452 user manual.pdf Chauvin Arnoux L452 user manual
Category
Technical document
File type
Adobe Acrobat
Download (4 MB)