Close

Are you visiting bsria.com from outside your region? Visit your regional site for more relevant services and pricing.

Pegasor Oy AQ indoor air quality meter

Contact Instrument Solutions

Overview

The Pegasor AQ is an easy to use indoor air quality meter that measures ultrafine particles (sizes greater than 0.01 micron and less than 2.5 micron), temperature, humidity and CO2 values. In addition to the internal datalogging capabilities the instrument can output measurements via a range of connectors including Bluetooth, Ethernet, analogue (V and mA) and optional 3G / 4G as well as WLAN.

Particulate measurement result can be shown as particle mass concentration (PM2.5), particle number concentration (PN) or active particle surface area concentration (PA). The instrument can be run for short periods off an internal battery and the particulate measuring range is 0.001 to 200 mg / m3.

Features

Key features

  • Measures ultrafine particles
  • Detects temperature, relative humidity and CO2 concentration
  • Self diagnostic procedure guarantees accurate continuous measurements
  • Fits easily to any indoor environment
  • Measured data can be read from the touch screen in graphical and numerical form
  • Measured data can be stored on hard disk or USB or transferred to and external device such as a cloud service, or printed out as separate document
  • Easy maintenance and long maintenance interval

NOTE: Instrument Solutions rental models include the temperature, humidity and CO2 probe option as well as GSM capabilities (SIM card is not provided).

Technical

Sensor type Diffusion charging operating principle

Measuring range 0.001 to 200 mg / m3

Maximum responsive particle size 0.01 to 2.5 μm

Particle size cutter PM2.5

Sampling mode Built in suction pump

Flow rate 3 l / min

Resolution 0.001 mg / m3 (with 1 min integration time) 0.1 % of reading

Sampling rate 1 / sec

Operation interface 108 mm color display touch screen

Time constant (logging interval) 1 to 300 s adjustable by user

Battery operation time 10h - 24h -100h (depending of the sampling rate)

Battery type Li-ion

AC power supply External power supply, 100 to 240 VAC, 50 / 60 Hz

Operating temperature range 0 to 50 oC

Storage temperature range - 20 to + 60 oC

Communication interface Ethernet with Modbus protocol, optionally wireless (3G/4G modem)

Automatic zero setting function Yes

Document printer Yes, with optional device

Service period 12 months, depending environment / use

Configurable to PM, PN or PA measurements Yes

Temperature and relative humidity measurements Yes, with Vaisala HUMICAP sensor

CO2 concentration measurement Yes, with Vaisala CARBOCAP sensor

Analog output mode 0 to 10 V or 4 to 20 mA

Analog output range Adjustable by user

Alarm output Yes, for different alarms

Alarm setting point range Adjustable

Built-in data storage 1 GB

Data storage type Flash memory

Storage interval 1 s to 1 h

Maximum data points > 1,000,000

Dimensions 150 mm x 200 mm x 300 mm (wxhxl)

Weight < 1.5 kg

GPS for location indicator Yes, optional

Instrument connections:

USB in, USB out Yes

Bluetooth Yes

WLAN Yes, optional

Ethernet Yes

3G / 4G Yes, optional

Analog (V and mA) Yes

Support documents

  • Download
  • Category
  • File type
  • Download
Download
Pegasor AQ indoor air quality meter user manual.pdf Pegasor AQ indoor air quality meter user manual
Category
Technical document
File type
Adobe Acrobat
Download (3 MB)
Download
Pegasor AQ indoor air quality meter quick set up guide.pdf Pegasor AQ indoor air quality meter quick set up guide
Category
Technical document
File type
Adobe Acrobat
Download (2 MB)
Download
Pegasor AQ indoor air quality meter data sheet.pdf Pegasor AQ indoor air quality meter data sheet
Category
Technical document
File type
Adobe Acrobat
Download (1 MB)
Download
Pegasor PPS plotter software user manual.pdf Pegasor PPS plotter software user manual
Category
Technical document
File type
Adobe Acrobat
Download (7 MB)

Common customer questions & answers

QQuestion Can I program the Pegasor AQ to perform a timed logging start?
AAnswer No. Once the unit has been switched on then it will continuously log and no programming is required.
QQuestion Can the Pegasor AQ monitor temperature and humidity remotely?
AAnswer No. The unit has its own attached temperature and humidity probe and no other sensors can be connected.