Designed to assess ventilation systems and measure pollutants in the context of indoor air quality monitoring, the Chauvin Arnoux CA1510 instrument is an accurate preliminary diagnostic indicator thanks to its alerts set at 1,000 ppm or 1,700 ppm of CO2. It operates from its internal batteries or can be run from a mains / micro USB charger.
Two communication modes are available: USB or Bluetooth. They can be used to configure the recordings, remote display on a PC or Android tablet and data display as a graph or as a table of values. They can also be used to export the data into Excel, to calculate the confinement index with selection of periods when people are present and to generate reports.
Features
- Recording capacity of up to 1 million values
- Logging intervals selectable from 1 minute to 2 hours
- USB interface (product recognised as a USB key) or Bluetooth for wireless use
- Software supplied as standard and 1510 application available with Android
- Locking of the instrument possible during recording (display & keyboard)
Technical
CO2 range 0 to 5,000 ppm
CO2 accuracy ± (50 ppm + 3 % of the value measured)
Temperature range - 10 °C to + 60 °C
Temperature accuracy ± 0.5 °C
Humidity range 5 to 95 % rh
Humidity accuracy ± 2 % rh
Battery life from new 1.5V AA/LR6 batteries you could get the following usage (rental kits may vary):
- In portable mode: 15 days
- In 1D 3D mode: 45 days
- In P_REC 10 minutes mode: 45 days
- In ECO mode: approximately 1 year